2000 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.
IEEE Transactions on Microwave Theory and Techniques
Volume 48 Number 11, November 2000
Page 1906.
Abstract:
A double-electrode applicator for localized thermal therapy is developed by breaking through the construction principle of the capacitive applicator. To make the beam of the electric field narrow, a ferrodielectric material is introduced as the sub-electrode. The optimum configuration of the double electrode and the material constant in each part of this applicator are investigated theoretically and experimentally. Good localized deep-heating characteristics are obtained at an operating frequency of 13.56 MHz and an output power 450 W by choosing a ferrodielectric material (BaTiO3) with a relative permittivity of 6000 as the sub-electrode.